An HPLC purity percentage is a share of ultraviolet peak area: how the material that gives a peak divides between the intended sequence and related impurities. Water, trifluoroacetate and other counter-ions, salts and UV-transparent excipients sit outside it, so a lot reported at 99 per cent can be far less than 99 per cent peptide by weight. Net peptide content is a separate measurement.
The answer: a share of peak area, not of mass
On a peptide certificate of analysis, HPLC purity is a share of peak area, not of mass. The sample is separated by reversed-phase liquid chromatography, ultraviolet absorbance is read at about 210 to 220 nm, where the peptide bond absorbs, and the main peak's area is divided by the total area of the peaks counted. Water, counter-ions such as trifluoroacetate, salts and UV-transparent excipients add weight without being counted in that total. A lot reported at 99 per cent can therefore be far less than 99 per cent peptide by weight, and the purity line cannot show how much peptide a vial holds against its label.
Bachem's published FAQ defines HPLC purity as the share of requested peptide among all material absorbing at 220 nm, and net peptide content separately, as peptidic material relative to counter-ions and residual water. Verbeke and colleagues, describing their method in 2015, calculated purity by area normalisation at 210 or 215 nm, applied the European Pharmacopoeia's reporting threshold of 0.1 per cent of the main peak area, and excluded solvent and system peaks.
Pages examined for this note state that more than 98 per cent purity means at least 98 per cent of the sample is the target peptide. None cited a source, none was found, and the measurements below contradict it. The claim is logged as untraced.
Area is not mass, and one peak can hold several
Reading area as amount assumes that every species absorbs alike. Kuipers and Gruppen measured molar extinction coefficients at 214 nm in 2007: against the peptide bond, tryptophan absorbed about 30 times as strongly, phenylalanine, tyrosine and histidine about six times, and proline within a chain about three times, while most other residues absorbed much less. An impurity that has lost an aromatic residue gives a smaller peak per milligram than its parent, so area per cent tracks mass per cent only as far as impurities absorb like the main peptide.
The figure also depends on what the method separates. Among the 98 peptides described below, Verbeke and colleagues found one whose supplier certificate stated at least 95.0 per cent from a method read at 220 nm; their own system, read at 210 nm, measured 80.6 per cent, and they judged the supplier's method to lack selectivity. Wang and colleagues reported in 2020 that related impurities in a synthetic glucagon could not all be separated from the main peak, and resolved them by mass. Zeng and colleagues, at an FDA laboratory, showed in 2015 that liquid chromatography with high-resolution mass spectrometry could quantify impurities in salmon calcitonin and two other peptide drugs even when they co-eluted, and detect them below 0.1 per cent of the active ingredient. One UV peak reported at 99 per cent can contain sequences that elute with it.
What the percentage leaves out
Mass balance puts numbers on the remainder. Wang and colleagues measured a commercial synthetic glucagon with Karl Fischer titration for water, ion chromatography for trifluoroacetate, inductively coupled plasma mass spectrometry for inorganic ions and LC-MS for nine related impurities. The manufacturer reported purity as 983.72 mg/g, and the authors' LC-UV gave more than 970 mg/g. Trifluoroacetate measured 103.03 mg/g, inorganic ions under 0.1 per cent, and glucagon 896.36 mg/g. The paper gives water as 50.2 mg/g in its results table but 0.502 mg/g in its mass-balance equation, and the final figure follows the smaller; the table's value would leave about 847 mg/g.
Melanson and colleagues assigned a value to a candidate certified reference material of angiotensin II in 2018, combining quantitative NMR, amino acid analysis by isotope dilution mass spectrometry, and mass balance. Trifluoroacetate made up nearly 25 per cent of the material by mass, and the assigned mass fraction of angiotensin II was 691 mg/g, plus or minus 9. Chen and colleagues reported in 2026 on sublancin, a peptide purified for use in feed additives: structure-related impurities were 0.4 per cent of the material, water 5.46 per cent and trifluoroacetate 15.48 per cent, and sublancin itself 78.60 per cent by mass balance and 77.58 per cent by amino acid analysis.
Material that gives no UV peak can weigh more still. Choules and colleagues reported in 2020 on two of three custom peptides bought from a major international manufacturer, with supplier HPLC-UV purities of 98.44 and 98.34 per cent. Quantitative proton NMR found mannitol, a sugar alcohol that is transparent in the ultraviolet and runs with the solvent front in generic reversed-phase methods, at 20 and 43 per cent by weight, and put the peptides at about 80 and 53 to 57 per cent.
Trifluoroacetate, left by synthesis and purification
Trifluoroacetic acid cleaves the chain from the resin in Fmoc solid-phase synthesis and pairs with basic groups during reversed-phase purification, so freeze-dried peptides are left as trifluoroacetate salts. Erckes and colleagues measured four angiotensin peptides in 2025. One counter-ion per positive charge predicted about 25 per cent trifluoroacetate by weight; measured content reached 35 per cent. One freeze-drying cycle in 10 mM hydrochloric acid brought it below the limits of quantification, chloride then made up as much as 10 per cent of weight, and the HPLC-UV purity of the first peptide stayed above 96 per cent.
The counter-ion is not inert in assays. Cornish and colleagues reported in 1999 that trifluoroacetate at 10 to 100 nanomolar reduced cell numbers and thymidine incorporation in fetal rat osteoblast cultures within 24 hours, with similar effects in articular chondrocytes and neonatal mouse calvariae. Trifluoroacetate salts of amylin, amylin-(1-8) and calcitonin gave consistently less proliferation than hydrochloride salts, enough to miss a proliferative effect or to record an antiproliferative one wrongly.
Exchange has its own costs. Roux and colleagues noted in 2008 that the classical route, repeated freeze-drying with excess hydrochloric acid, means working below pH 1, which can degrade peptides. With the octapeptide lanreotide, reversed-phase HPLC and an ion-exchange resin gave partial to almost complete exchange to acetate, and a deprotonation and reprotonation cycle removed trifluoroacetate entirely.
Net peptide content, measured separately
Net peptide content answers the other question: how much of the weighed powder is peptide. Water and the usual counter-ions carry no nitrogen, so Bachem's FAQ derives it from nitrogen by elemental analysis, or from amino acid analysis; quantitative NMR and mass balance also measure it. Because peptidic impurities count as peptide in that definition, the intended sequence is the purity share of the net content. Vemuri compared four methods on lyophilised thymalfasin in 2005: amino acid analysis was highly variable in one laboratory and precise in another, HPLC results depended on the laboratory, and elemental analysis gave coefficients of variation under 2 per cent, the method the author favoured.
Supplier handbooks give ranges without data. AmbioPharm's FAQ puts net peptide content in most peptides at 60 to 90 per cent of gross weight; LifeTein's guide states that a lyophilised peptide may carry 10 to 70 per cent non-peptide material by weight. Neither cites a dataset, their lower bounds disagree, and both are recorded here as guidance. The measured peptide fractions above run from 530 to 896 mg/g.
The two figures can sit on one certificate without moving together. On one tesamorelin lot certificate that lists both figures, for a lot tested on 4 September 2026, HPLC purity is 99.90 per cent against a specification of at least 95 per cent, and a separate line gives net content as 10.67 mg in a vial labelled 10 mg. The first is a ratio of peak areas; the second is a mass, and on that lot it came out above the label. Neither can be calculated from the other.
What tests of purchased peptides found
Verbeke and colleagues ordered 98 quorum-sensing peptides from two suppliers at a requested purity of at least 95.0 per cent, and each arrived with a certificate stating that. Their own analysis found 43 that met it. None of the nine cyclic peptides exceeded 90 per cent, a level the suppliers then said they could not reach, and the main peak of one peptide was a different structure, consistent with disulfide cyclisation. Of 61 impurities assigned structures, 74 per cent lacked at least one residue of the intended sequence.
Ashraf and colleagues, reporting in 2024, measured vials bought online against their labels. Of six test purchases from illegal online pharmacies, three vials of lyophilised semaglutide arrived, presented as research products, with 99 per cent purity claimed on the label or the seller's website. LC-MS found one peptide signal in each and semaglutide 28.56 to 38.69 per cent above the labelled 1 to 3 mg, yet semaglutide made up only 7.70 to 14.37 per cent of the powder, which weighed 14.9 to 50.1 mg. The rest was not identified, and endotoxin was present in all three.
Janvier and colleagues screened ten falsified peptide drugs from three suspected illegal internet pharmacies on the Belgian market in 2018, and reported wide variation in drug per unit, purity of 5 to 75 per cent among the cysteine-containing peptides, and arsenic and lead in some samples. PubMed searches run on 10 November 2026 found no study that measured trifluoroacetate or water in research-labelled peptides sold online.
What traces to a source, and what does not
The claim that more than 98 per cent purity means a vial is at least 98 per cent peptide traced to nothing: the pages making it cited no source, a PubMed search pairing HPLC or chromatographic purity with peptide content or mass fraction returned no records, and the glucagon, mannitol and semaglutide measurements contradict it. The claim that research grade means 98 per cent or higher also traced to nothing. No pharmacopoeial or regulatory definition of research grade was located, the pages examined gave 95, 98, or 95 to 99 per cent, a PubMed search for research grade with peptide purity returned one record that defines no grade, and Verbeke and colleagues wrote that peptides used in research range from 50 to more than 95 per cent purity. Both claims are logged as untraced.
The measured record is narrower. HPLC purity reported the UV-absorbing share in the main peak, trifluoroacetate made up about 10 to 35 per cent of the mass of the peptides measured for it, and a lot's peptide content was known only where it was measured.
Disclosure. Vialog is published by the team behind Disguised Alpha, which sells the product linked in this note. The note follows the same editorial policy as every other page on this site.
References
- Bachem. FAQ: frequently asked questions (Knowledge Center). Supplier FAQ; no author shown; page metadata: published 9 March 2021, modified 15 October 2026. Entries read: what HPLC purity means (percentage of requested peptide relative to all material absorbing at 220 nm), 'assay' as nitrogen content by elemental analysis, gross weight, net peptide content (peptidic material, including peptidic impurities, relative to counter-ions and residual water; by elemental analysis or amino acid analysis). No study, dataset or reference cited. Guidance, not a study. No PMID. Retrieved 10 November 2026. View on bachem.com
- Verbeke F, Wynendaele E, Braet S, D'Hondt M, De Spiegeleer B. Quality evaluation of synthetic quorum sensing peptides used in R&D. J Pharm Anal. 2015;5(3):169-181. doi:10.1016/j.jpha.2014.12.002. PMID 29403929. Method (area normalisation at 210 and 215 nm; Ph. Eur. reporting threshold 0.1 per cent and identification threshold 0.5 per cent of main peak area; solvent and system peaks excluded), the 98 peptides, 43 meeting 95.0 per cent, the nine cyclic peptides, peptide ID76 (supplier at least 95.0 per cent, in-house 80.6 per cent), ID25, the 84 impurities and the 50 to more than 95 per cent statement read in the full text, PMC5762210. View on pubmed.ncbi.nlm.nih.gov
- Kuipers BJ, Gruppen H. Prediction of molar extinction coefficients of proteins and peptides using UV absorption of the constituent amino acids at 214 nm to enable quantitative reverse phase high-performance liquid chromatography-mass spectrometry analysis. J Agric Food Chem. 2007;55(14):5445-51. doi:10.1021/jf070337l. PMID 17539659 View on pubmed.ncbi.nlm.nih.gov
- Wang X, Zhang F, Li H, Xiao P, Su F, Xu B, Sun W, Song D. Purity determination of synthetic glucagon using a mass balance approach. Sci Rep. 2020;10(1):4423. doi:10.1038/s41598-020-61109-9. PMID 32157163. Manufacturer purity 983.72 mg/g, in-house LC-UV over 970 mg/g, trifluoroacetate 103.03 mg/g (Table 1), water 50.2 mg/g (Table 1) against 0.502 mg/g in the mass-balance equation, inorganic ions under 0.1 per cent, nine impurities by LC-MS and the final 896.36 mg/g read in the full text, PMC7064584. View on pubmed.ncbi.nlm.nih.gov
- Zeng K, Geerlof-Vidavisky I, Gucinski A, Jiang X, Boyne MT. Liquid Chromatography-High Resolution Mass Spectrometry for Peptide Drug Quality Control. AAPS J. 2015;17(3):643-51. doi:10.1208/s12248-015-9730-z. PMID 25716148 View on pubmed.ncbi.nlm.nih.gov
- Melanson JE, Thibeault MP, Stocks BB, Leek DM, McRae G, Meija J. Purity assignment for peptide certified reference materials by combining qNMR and LC-MS/MS amino acid analysis results: application to angiotensin II. Anal Bioanal Chem. 2018;410(26):6719-6731. doi:10.1007/s00216-018-1272-7. PMID 30143839 View on pubmed.ncbi.nlm.nih.gov
- Chen W, Yan J, Yang M, Wang M, Zhou J, Li F, Li L. SI-traceable purity assignment for peptide sublancin using mass balance approach and isotope dilution mass spectrometry based on amino acid analysis. J Chromatogr B Analyt Technol Biomed Life Sci. 2026;1277:125038. doi:10.1016/j.jchromb.2026.125038. PMID 41916078 View on pubmed.ncbi.nlm.nih.gov
- Choules MP, Bisson J, Simmler C, McAlpine JB, Giancaspro G, Bzhelyansky A, Niemitz M, Pauli GF. NMR reveals an undeclared constituent in custom synthetic peptides. J Pharm Biomed Anal. 2020;178:112915. doi:10.1016/j.jpba.2019.112915. PMID 31671336. Supplier HPLC-UV purities (98.44 and 98.34 per cent) and integral-based qHNMR purities (80.2 and 53.0 per cent) read in the full text, PMC6913887. View on pubmed.ncbi.nlm.nih.gov
- Erckes V, Streuli A, Chamera Rendueles L, Krämer SD, Steuer C. Towards a Consensus for the Analysis and Exchange of TFA as a Counterion in Synthetic Peptides and Its Influence on Membrane Permeation. Pharmaceuticals (Basel). 2025;18(8):1163. doi:10.3390/ph18081163. PMID 40872554. Predicted (about 25 per cent) and measured (up to 35 per cent) trifluoroacetate, chloride up to 10 per cent, the 10 mM HCl exchange and the purity above 96 per cent read in the full text, PMC12389442. View on pubmed.ncbi.nlm.nih.gov
- Cornish J, Callon KE, Lin CQ, Xiao CL, Mulvey TB, Cooper GJ, Reid IR. Trifluoroacetate, a contaminant in purified proteins, inhibits proliferation of osteoblasts and chondrocytes. Am J Physiol. 1999;277(5):E779-83. doi:10.1152/ajpendo.1999.277.5.E779. PMID 10567002 View on pubmed.ncbi.nlm.nih.gov
- Roux S, Zékri E, Rousseau B, Paternostre M, Cintrat JC, Fay N. Elimination and exchange of trifluoroacetate counter-ion from cationic peptides: a critical evaluation of different approaches. J Pept Sci. 2008;14(3):354-9. doi:10.1002/psc.951. PMID 18035848 View on pubmed.ncbi.nlm.nih.gov
- Vemuri S. Comparison of assays for determination of peptide content for lyophilized thymalfasin. J Pept Res. 2005;65(4):433-9. doi:10.1111/j.1399-3011.2005.00225.x. PMID 15813890 View on pubmed.ncbi.nlm.nih.gov
- AmbioPharm. What is net peptide content? Supplier FAQ; no author shown; page metadata: published 14 March 2019, modified 15 September 2022. States that net peptide content is usually 60 to 90 per cent of gross weight in most peptides, determined by elemental analysis, amino acid analysis or UV spectrophotometry. No study, dataset or reference cited. Guidance, not a study. No PMID. Retrieved 10 November 2026. View on ambiopharm.com
- LifeTein. How to calculate the peptide concentration. Supplier guide; no author or date shown. States that a lyophilised peptide sample may contain 10 to 70 per cent non-peptide material by weight, depending on sequence, purification and formulation. No study, dataset or reference cited. Guidance, not a study. No PMID. Retrieved 10 November 2026. View on lifetein.com
- Ashraf AR, Mackey TK, Vida RG, Kulcsár G, Schmidt J, Balázs O, Domián BM, Li J, Csákó I, Fittler A. Multifactor Quality and Safety Analysis of Semaglutide Products Sold by Online Sellers Without a Prescription: Market Surveillance, Content Analysis, and Product Purchase Evaluation Study. J Med Internet Res. 2024;26:e65440. doi:10.2196/65440. PMID 39509151. Powder weights, labelled and measured semaglutide amounts, labelled and measured purity (Table 4), the single LC-MS signal and the research-use labelling read in the full text, PMC11582493. View on pubmed.ncbi.nlm.nih.gov
- Janvier S, Cheyns K, Canfyn M, Goscinny S, De Spiegeleer B, Vanhee C, Deconinck E. Impurity profiling of the most frequently encountered falsified polypeptide drugs on the Belgian market. Talanta. 2018;188:795-807. doi:10.1016/j.talanta.2018.06.023. PMID 30029448 View on pubmed.ncbi.nlm.nih.gov